MTS Series 793 Application User Manual

Page 396

Advertising
background image

MTS MultiPurpose TestWare®

Temperature Control Process

External Control Process Descriptions

396

General tab

Channels tab

Displays the channels available to this process. Use the arrow keys to move
desired channels from the Available list to the Included list.

Monitor Signal

Specifies the temperature feedback signal associated with the monitor feature.

This feedback comes from a second temperature sensor in the enclosure, which

is conditioned by its own conditioner. This feedback is independent of the
feedback used by the external temperature controller to achieve closed-loop

control.

Monitor Tolerance

Specifies the amount the monitor temperature feedback may vary from the
specified End Level during the Dwell Period.

If you use the monitor feature, the Temperature Control process monitors the
feedback associated with the Control Tolerance and the feedback associated

with the Monitor Tolerance concurrently. If either feedback strays outside of

its tolerance range the dwell period starts over.

Command Tab (part 2 of 2)

I

TEM

D

ESCRIPTION

General Tab

I

TEM

D

ESCRIPTION

Name

Specifies the process name displayed on the Procedure or Group process

window.

Process Enabled

Enables the process. (Clear this box to disable the process.)

Execute Process

Specifies the number of times the process will be run.

Counter Type

Enables a process counter.

Options are None, Transient, or Fixed. Transient counters are displayed only
when the process is active. Fixed counters are displayed throughout the test.

Counter Label

Names the counter.

When you run the test, this name appears under Sequence Counters on the

MPT control panel.

Note

If this box is blank, the process name is used as the counter label.