Set Controller Value Process

MTS MultiPurpose TestWare®

Other Process Descriptions

433

General tab

How to Define the Set Controller Value Process

Small calculation parameter changes can produce large control-loop
changes, resulting in violent actuator movement.

Violent actuator movement can result in injury and equipment damage.

Use care when changing calculation parameters.

Note

To use this process, you must have calculation parameters and/or
calculation variables defined in the station configuration file. You define
calculation parameters and calculation variables with the Equation Editor
in the Station Manager application.

1. Drag the Set Controller Value process icon

from the Process Palette

to a Group or Procedure table.

2. Click the Set Controller Value process icon in the Group or Procedure

table.

General Tab

I

TEM

D

ESCRIPTION

Name

Specifies the process name displayed in the Procedure or Group process window.

Process Enabled

Enables the process. (Clear this box to disable the process.)

Execute Process

Specifies the number of times the process will run.

Counter Type

Enables a process counter.

Options are None, Transient, or Fixed. Transient counters are displayed only when

the process is active. Fixed counters are displayed throughout the test.

Counter Label

Names the counter.

When you run the test, this name appears under Sequence Counters on the MPT
control panel.

If this box is blank, the process name is used as the counter label.

WARNING