Index 2MF

Users Guide

D-2

Dynamic Test

Test Format........................................................... Photo detector/diode response to both the red and infrared light

generated by the probe when pulsed by an internal test signal.

Test Signal.......................................................... Pulsed between the two LEDs, constant current level at 1.0 mA
Test Results........................................................ Nominal range of 0 to 20,000

Checksum.............................................................. The sum of all locations in the program chip. For service only.

Note

The above specifications are for pre-programmed makes only, 20 % pulse

amplitude SpO2. Results for Presets, custom makes, and various pulse

amplitudes are unspecified and depend on the algorithm of the specific

pulse oximeter being tested.