Fluke Biomedical 2MF Index User Manual

Page 107

Advertising
background image

Appendices

Computer Control

E

E-7

The Baud Rate in the Simulator must be set to match the LD.BAT Baud Rate setting as

shown in Table E-4.

Table E-4. Baud Rate Settings

LD.BAT Setting

Baud Rate

30 300

60 600

12 1200

24 2400

96 9600

Note

When using the LD.BAT program to download waveform shapes, only

download one waveform shape or R-Curve file at a time. Do not string

several shapes or R-Curves into one file for download.

Player Mode

Player mode permits files of red and infrared readings, taken off an oximeter, to be

"played back" through the Simulator in real time. These are long-duration files (player

files) of real physiological behavior. These files are only available from select pulse

oximeter manufacturers. Fluke Biomedical does not supply them.
Player file data is uploaded through the Simulator’s serial port and loaded directly into

the red AC and infrared AC digital-to-analog converters (DACs). The output waves of

the DACs are controlled by the amplitude values and update rate of the serial port. Data

sampling rate and playback rate will be equal. Data will nominally be set at 9600 baud

N,8,1. However, this rate may be adjusted to control playback rate in terms of

points/second. The N,8,1 must not change.
Once a player file (named with a .BIN extenstion) is obtained, a small batch file needs to

be created and placed in the same directory (or folder). The batch file should consist of

the following commands:

mode com1:96,n,8,1

:label

copy /b %1 com1:

goto label

Note

The /b in the batch file forces DOS to handle the file as a binary file.

To use the Player Mode:
1. Send [player] to the Simulator via computer control.
2. In DOS, go to the directory that has the two files and issue the following command:

LD filename (example: LD 8sec.bin), and press ENTER.

Advertising
This manual is related to the following products: