Index 2MF

Users Guide

G-8

In some cases, it may be advantageous to not use the floating diode source. In these

cases, the Simulator provides a ground pin at pin 9, and both positive and negative

simulation voltage signals on pins 7 and 8. These may be converted to current signals by

appropriate scaling resistors as noted in Figure G-5.
Normally, oximeter shield pins should not be tied to the Simulator ground, but there may

be exceptions. Again, only a thorough study and understanding of the oximeter interface

circuits can answer these questions.

EXAMPLE: OHMEDA HARDWIRE CABLE

INDEX 2

PIN

OHMEDA PROBE

PIN

TO OXIMETER

IR CATHODE

R CATHODE

RED ANODE

15.8

ID RES

ID RES COMMON

DET. ANODE

SHLD/DET RET

T

1

2

3

4

5

6

7

8

9

1

2

3

4

5

6

7

8

9

X2

56K
1%

200K, 1%

100K, 1%

esl163.eps

Figure G-5. Example: Ohmeda Hardware Cable