When the alarm sounds, press the softkey STOP.

esl071.eps

The timer ceases counting and the elapsed time (in seconds) displays on the screen.

Note

You can connect the Simulator to any printer with a serial or parallel
Centronics-style port or to any IBM-compatible PC, including laptops and
notebooks.

5. To print the results of the test for review or archiving, press PRT.

esl072.eps

Selecting PRT sends results to the RS232 and Centronics ports. With a printer or
computer attached, the results should look something like this example:

Asystole Response Time

---------------------

24.0 sec

For information on how to connect to other systems or peripherals, see “Appendix A:
Printing & Data Transfer”.

6. When you are done, press the softkey esc to return to the Limits menu.