Operation – Fluke Biomedical 2MF Index User Manual

Page 32

Advertising
background image

Index 2MF

Users Manual

5-4

Note

The Simulator accuracy specifications are not valid for user defined custom
makes. The Simulator can provide specified repeatability of

±

1 standard

deviation.

Table 5-1. Oximeters with Technology Options

Technology Options

Typical Oximeters

Normal Intensity

Nellcor, BCI

Low Intensity

Hewlett-Packard, Ohmeda, Invivo

High Intensity

Datex

Very High Intensity

None known

Table 5-2. Oximeters with R-Curve Parameters

Curve No.

Brand

HiOx Level

Oxy @ R = 0.7

LoOx Level

Oxy @ R = 1.0

1 BCI

93

82/83

2 Criticare

94

80/79

3 Datascope

94

85

4 Datex

95

84

5 Hewlett-Packard

96

85

6 Nellcor

93

80

7 Hihon

Kohden

94

84

8 Novametrix

96

87/88

9 Ohmeda

95

86/85

10 Respironics

93

80

Once a technology and a curve have been determined, save the custom make for future
use.

Operation

To access the custom menu, press

CUST

from the main menu.

Chose the slot to customize by pressing

+CST

to cycle through the available names.

Custom #1 through #5 are the default factory names. To change the name press

NAME

.

There are five available slots for saving custom makes. Press

+

or

to change the

alphabet. Press

to move the character. Press

SAVE

to save the name and the results.

Press TEC to test for technology type.

Select the appropriate technology type #1 - #4. Technology types offer various light level
thresholds. Refer to Table 5-1 for a list of technology light levels. The technology to use
may already be known based on the manufacturer’s information. If not, continue for
process of elimination testing.

Press

HiOx

to test oximeter stabuility at the high oxygen setting.