Using Test Programs

The AUTO Menus

8

8-3

Table 8-1. Auto Test Parameters (cont.)

Name Description

Make

Set the make of machine being tested, such as Nellcor, HP,
Criticare, Ohmeda, Invivo Research, and so on.

O2 Alarm Test

Set the Simulator to either test or bypass the SpO2 limits test.

BPM Alarm Test

Set the Simulator to either test or bypass the BPM units test. If

Yes

, an alarm will sound at the conclusion of testing the pulse

rate.

Pulse Amplitude Test

Set the Simulator to either test or bypass the amplitude test.

Motion Test

Set the Simulator to test the oximeter for patient motion.

Use Presets

Set the Simulator to allow for the selection of preset patient
conditions (geriatric, obese, normal, etc.) at test time.

Accessing the Autosequences Menu (AUTO)

To access the Autosequences menu, press

more

on Main Menu 1 and then press AUTO

on Main Menu 2. See the section “Creating a Custom Test Program” for the additional
parameters available under each option.

fgg181.eps

After you press

AUTO

, the following options become available:

esl075.eps

The Autosequences menu provides the following options:

PROG

to define and edit programs

RUN

to start an autosequence

DEFS

to restore defaults

esc

to return to prior menus