Fluke Biomedical 2MF Index User Manual

Page 44

Advertising
background image

Index 2MF

Users Manual

7-2

esl049.eps

The display changes to present these options:

esl051.eps

When setting new oxygen limits, the Simulator displays the last setting used (or 96%
if not previously changed) and resets the time to 00.0 seconds.

2. Press the softkeys

+

or

-

to start the timer, and the screen displays the “Timing!”

message.

esl052.eps

Pressing the softkey

+

(plus) will increase by the step amount set previously

using the Simulations menu without affecting the timing function.

Pressing the softkey

(minus) will decrease by the step amount without

affecting the timing function.

Note

Continue pressing

+

or

-

until the O2 percent you want to test for is

displayed. Each press pf the softkey

+

or

resets the timer internally and

restarts the timing.

3. If the O2 setting is outside the oximeter's alarm limit, the alarm will sound on the

pulse oximeter being tested after a period of time. Press STOP immediately. This
stops the timer, and displays the “elapsed time to alarm” for the specified simulation
level.

esl053.eps

4. To print the results of the alarm, press PRT.