Index 2MF

Users Guide

E-4

Command Description Values

Example

MFREQ

[1]

nn.nnn

Sets the artificial light artifact frequency in Hertz.
Internal resolution at 60 Hz is about 0.2 Hz. To
let the user know the effects of input rounding,
the rounded value is output on the RS-232. For
example, if the user enters [mfreq 60.1], the
result will be 60.1882. Entry resolution is 0.001;
however, the resolution of the value stored and
used is near 0.2 Hz at 60 Hz.

45.000 to
75.000

[MFREQ 60.1]

MOXY nnn

Sets the oxygen value of the non-arterial wave,
tap, or shiver. Leading zeros are not required.

0 to 100

[MOXY 75]

NOMOTION Turns off tap and shiver motion waves.

- -

[NOMOTION]

OXY

Sets the SpO2 percent for the selected make.

35 to 100 [OXY96]

PLAYER

Turns on the player mode.

- -

[PLAYER]

PRESETS
nn

Selects a specific preset, where nn ranges from
00 to 23. A range of 00 to 23 for nn corresponds
to preset motion selection shown in the presets
in Table 6-2.

00 to 23

[PRESETS 05]

PROBE

Performs and returns the electrical probe test
results. Displays tests by line.

- -

[PROBE]

Q

Quits the communications session and returns
Index 2MF to Index 2MF control.

- -

[Q]

RAC

Sets the red light AC amplitude peak-to-peak
attenuation for the pulse oximeter under test.

0 to 4095 [RAC4000]

RCURVE

Sends a new R-curve to Index 2MF.

various

[RCURVE, 0,
NELLCOR, 2.00,
1.999, 1.899, 1.799,
1.699 …]

RDC

Red transmissivity.

0 to 4095 [RDC1000]

SETTINGS

Retrieves the current Index 2MF settings in this
format:
Index 2MF Settings:
-------------------
oxy displayed after set by [oxy NN] command
BPM=75

Note: Modifying the following 4 variables inhibits
oxy display until the next [oxy NN] command.

RAC = 610
IRAC = 1000
RDC = 1024

IRAC = 1024
Light Flash: Unit Under Test
AC = 100 %
AMB = 0 %
Make = Nellcor.

- -

[SETTINGS]