Index 2MF

Users Guide

4-2

fgg010.eps

Selecting SIM opens the menu displaying the following:

MAN to run manual simulations

PRESET to run factory-defined simulations

AMB to set ambient light levels

CUST to create custom makes

esc softkey to return to Main Menu 1

LMTS lets you set the alarm limits for oxygen, beats per minute, pulse amplitude, and
lets you set up a timer for asystole.

fgg011.eps

Selecting LMTS opens the menu displaying the following:

O2 to set up Sp02 alarm limits level and response time

BPM to set up rate alarm limits and response time

AMP to select pulse amplitude limits

ASYS to set up asystole

PRBE offers the electrical probe test menu to analyze oximeter probe integrity. Tests
include LED voltages, photo diode results, and pin-to-pin resistance.

fgg012.eps

Selecting PRBE tests attached probes and displays the following options:

LEDS to view the red and infrared LEDs

PHTO to view photo diode results

RES to view pin-to-pin resistance

REDO (requires an attached probe)

esc softkey to return to Main Menu 1