This Appendix provides information on the Simulator’s RS232 and printer ports,

including how to configure the port on your printer or computer.
The Simulator provides a male D (9-pin) bi-directional serial port located at the back side

of the unit, for the transfer of data to a computer or printer, and for computer control of

the Simulator. The Data Computer Equipment (DCE) wiring configuration is shown

Figure A-1.

Index 2MF RS232 Configuration

Fluke Biomedical offers a bi-directional serial null modem cable for connect the

Simulator to your computer (IBM compatible) or serial printer. Please specify Fluke

Biomedical Part Number 2238626 when ordering.

1- C-Lock
2- TX
3- RX
4- DTR/DSR
5- Ground

6- DTR/DSR
7- RTS/CTS
8- RTS/CTS
9- Unused

esl165.eps

Figure A-1. Male D Cable End

C-Lock

Pin 1 of the serial connector provides the Simulator’s synchronized pulse. This pulse is

normally in phase with the start of each 64-point pleth wave. The phase of the sync pulse

with regard to the wave can be varied using the [CPHASE nn] command, where nn is a

number in the range of 0 to 63. The pulse is a negative going attenuated logic level. For

more information, refer to “Appendix E: Computer Control”.