Testing the Pulse Oximeter Limits

Simulating Pulse Amplitude

7

7-5

esl060.eps

The results will be sent to the RS232 and Centronics port.

Note

Connect the Simulator to any printer with a serial port or parallel
Centronics port, or to any IBM-compatible PC, including laptops and
notebooks. For information on how to connect to other systems or
peripherals, see “Appendix A: Printing & Data Transfer”.

If a printer or computer is attached, you will see results that look something like this
example:

Rate Alarm Response Time
-------------------------
10.0 sec

Press esc to return to the Limits menu.

esl061.eps

Simulating Pulse Amplitude

The peak-to-peak amplitude of the simulated blood pressure wave can be increased or
decreased. Decreasing amplitude corresponds to a weakening pulse. You can decrease
amplitude to find where the oximeter fails to find the pulse.

You can quickly set a pulse amplitude limit.

1. To access the pulse amplitude alarm menu, press LMTS from Main Menu 1, and then

AMP from the Limits menu:

esl062.eps

2. The Pulse Amplitude menu displays in the LCD screen: