This is the equation used to obtain the R-values used in the Simulator, and presented on
the LCD screen.

Although the equation eliminates the effect of changes in the UUT flash amplitude on the
R-value, bulk attenuation of an actual finger is an important characteristic. By varying
rdc or irdc values and determining how large a range of variation is “not seen” by the
oximeter being tested, you can get a good idea of the oximeter's electronics' dynamic
range and its processing capability. This is independent of finger thickness and
pigmentation.

Note

By varying the ac term, you can determine the oximeter's ability to track
and hold a pleth wave as it weakens to zero, and then re-acquire the wave
as it strengthens.

Accessing Manufacturers' Tests

Access the manufacturers' test settings with the LCD screen.

Press the softkey OPTS on Main Menu 2.

fgg181.eps

The Advanced Options menu displays in the LCD with two menu options in addition to
the escape softkey.

esl144.eps

The following options become available on the Advanced Options menu:

SRC Sets the pulse signal source

RVAL Sets the R-value

esc

Returns you to the Options Menu

Setting the Signal Source

1. Press the SRC softkey on the Advanced Options menu:

esl145.eps