Setting and changing simulation factors, About simulations and presets, Setting simulations – Fluke Biomedical 2MF Index User Manual

Page 35: The simulations menu, Chapter 6

Advertising
background image

6-1

Chapter 6

Setting and Changing Simulation Factors

About Simulations and Presets

One of the main functions of the Simulator is to test the performance of a pulse oximeter
for any possible patient condition. The Simulator allows you to set any combination of
blood-oxygen levels, heart rates, and pulse amplitude, as well as simulate a virtual range
of medical physical conditions and set an ambient light artifact.

Setting this group of default factors allows you to test an oximeter and determine its
weaknesses, if any. The Simulator will display and optionally print areas in your pulse
oximeter performance that may be unacceptable, and document correct performance.

The simulations set here are used as the defaults when testing the pulse oximeter.

Setting Simulations

To access the Simulations Menu, press the softkey SIM from the Main Menu.

fgg010.eps

The Simulations Menu

The Simulations menu appears in the LCD with these choices:

MAN to manually set the O2 and BPM

PRESET to set a patient’s health condition

AMB to set an ambient light artifact

STEP to change step size for saturation and BPM

esc to return to previous menu (Main Menu 1)