Appendix B

Error Messages and Corrective

Measures

Introduction

This Appendix lists all of the error messages or beeps that may appear on the Simulator's

LCD screen during normal operation. Solutions are given for each message.

Messages

If the battery has run down, the following message appears:

Warning! Low Battery

Although the battery included with the Simulator can run continuously for up to 8 hours,

it does need regular recharging, especially after long periods of disuse. In addition, the

battery can only be recharged a certain number of times, after which it needs to be

replaced.
To recharge the battery, plug the battery charger that came with the Simulator into the

Simulator charger jack and plug the battery charger into a wall socket, using the proper

line cord.

Note

To fully recharge the battery, allow the battery charging at least 12 hours.

Disconnecting the battery charger and using the battery prior to the full 12

hour charge time will result in an incomplete charge. The Simulator may

only work for a very short time under such circumstances.

Beep Indications

Three beeps
You have pressed a button that is not activated for the particular screen or function. Try

again.
Continuous beeps
Continuous beeps warn of a low battery. See the recharging instructions above.