Appendices

Computer Control

E

E-5

Table E-1. Valid Commands (cont.)

Command Description Values

Example

SHAPE

Indicates that the next 64 numbers represent a
pleth wave shape.

- -

[SHAPE1.000, 3.9,
4.9,...]

SHIVER

Causes shiver artifact, a motion wave of
6.0 Hz.

- -

[SHIVER]

SRCF

Selects DC or fixed signal source level.

- -

[SCRF]

SRCU

Selects the test unit's signal source level.

- -

[SCRU]

TAP

Causes tapping artifact, a motion wave of
3.5 Hz

- -

[TAP]

UPLETH

Uses user's pleth waveform.

- -

[UPLETH]

VER

Displays the Index 2MF version number

- -

[VER]

[1] For values less than 1, it is necessary to enter a leading zero. For example, 0.1 must be entered instead of

.1. A .1 entry will be interpreted as a signal strength of 1. Out of range entries default to six, which is a
normal level. In reality, signal strengths over 6 % always correspond to saturation in the 90s therefore the
80 % limit should be of no physiological consequence.

R-CURVE

The following is a sample of the format. Refer to Table E-2 for descriptions.

[ RCURVE>3:123456789:xxx:YYY:zzz,zzz,.........zzz,]

Table E-2. R-CURVE Format

Format Description

[ RCURVE> Identifies this to computer control as an R-curve data stream.

3:

The curve number. Range is 0 to 5. There are six total downloadable R-curves.
The colon is a field separator.

Pulse oximeters use different intensities of light. R-curves 0 to 3 are reserved for
oximeters that use normal intensities of light. R-curve #4 is reserved for pulse
oximeters that use higher intensities of light; R-curve #5 is for pulse oximeters that
use lower light intensities.

123456789

This field can be any combination of up to nine alpha-numerics or space
characters. NOTHING ELSE! Typically, it will be an oximeter brand name.

xxx

The irdc value for this oximeter. Range is 0 to 1000, divided by 10 and rounded.
For example, for 972 , use 097 (1000 becomes 100, and 17 becomes 020).
Leading zeros are required!

YYY

Rdc value for this oximeter. Same format as irdc.

: Field

separator

zzz

R-values in the range 0 to 250. These will be multiplied by .01 inside the Simulator
to be in the range 0 to 2.50. Leading zeros are required. Commas are required. No
spaces are allowed. A comma after the last value is required. 51 values are
required, corresponding to 50 thru 100 percent oxy, read left to right.

]

Right bracket identifies the end of the data stream.