E. computer control, Computer control, Appendix e – Fluke Biomedical 2MF Index User Manual

Page 101: Introduction

Advertising
background image

E-1

Appendix E

Computer Control

Introduction

You can set up a program to communicate with and control the Simulator via its

RS232 port. To control the Simulator by computer:
1. Switch the Simulator power on using the slide switch. The LCD sequentially

displays the following screens for about one second each before ending on the Main

Menu 1 screen:

Bio-Tek Instruments Inc. and model number and version

Software version number and copyright years

Make of last selected test equipment

Main Menu 1 screen

esl009.eps

2. Connect the Simulator to the computer by connecting the RS232 cable to both the the

Simulator and the computer’s ports. Use a serial or null-modem cable. The RS232

port is a male DB 9-pin port.

Note

See “Appendix A: Printing & Data Transfer” for information on how to

configure the Simulator's RS232 port to match the speed, bit settings, and

parity of your computer system's serial port.

3. To initiate computer control, send a left bracket ([) from your computer to the

Simulator.