Appendix g, Introduction – Fluke Biomedical 2MF Index User Manual

Page 113

Advertising
background image

G-1

Appendix G

Custom Probe Test Cables and Electrical

Simulation Cables

Introduction

This Appendix explains how to make custom probe adapter cables and electrical

simulation cables to be used with your Index 2MFE.

Making Probe Test Cables

Index 2MFE can be used to test probes using custom probe adapter cables.
Figure G-1 shows a typical oximeter probe. It consists of two LEDs and one photodiode.

In addition, it may also contain an LED spectrum characterizing resistor or two. Some of

the wires may be shielded.
Typical probe failure modes are:

Wires open or shorted to other wires

Diodes open or shorted

Resistors, if present, open or shorted to other leads

Leads shorted to shields and/or other leads

Advertising