Setting the 02 level, Setting the pulse rate – Fluke Biomedical 2MF Index User Manual

Page 59

Advertising
background image

Using Test Programs

Creating a Custom Test Program (Autosequence)

8

8-9

esl094.eps

3. Press SEL to change the Include Actuals: response from YES to NO as needed.

4. Move to the next step in the process by pressing the softkey ADV.

Setting the 02 Level

You can set up to ten different oxygen levels and up to ten different pulse rates for the
simulation autosequences.

Note

In each screen, the softkey

ADV

moves ahead to the next settable option.

Use the

NEXT

softkey to simply advance through the O2 and rate level

settings for each test.

1. After selecting a test, set the oxygen level for test number one:

esl095.eps

2. Press

+

or

-

to adjust the simulated oxygen level up or down for the first O2 test

(1% O2 per button push). You can set up to ten O2 tests. Pressing NEXT at Rate

Setting #10 brings you back to O2 Setting #1.

esl096.eps

3. Press NEXT

to set the simulated pulse rate.

Note

If you do not plan to test for O2, press ADV to bypass these settings.

Setting the Pulse Rate

You can set up to ten different pulse rates for the Simulator to use during a single
autosequence program.

1. Again, use

+

and/or

-

to adjust the BPM up or down for the first pulse rate test.