Index 2MF

Users Manual

7-6

esl063.eps

3. To raise the Pulse Amp percentage (AC factors) in the simulated blood condition,

press the softkey

+

:

esl064.eps

The amplitude will increase in increments of 1% with every softkey press.

4. To lower the percentage of AC factors in the simulated blood condition, press the

softkey:

esl065.eps

The amplitude will decrease by 1% with every softkey press down to 1 %. From 1 %
to 0.1 % the value will decrease by 0.1 %. Below 0.1 %, the value will decrease by
0.025 % per button press. Inputs below 6 % are rounded internally to the nearest
0.024 %, and above 6 % are rounded to the nearest 0.072 %. The displayed value is
the rounded value nearest the value displayed per the preceding increment/decrement
rules. If the pulse amplitude rate exceeds the oximeter alarm rate, the alarm will
sound on the pulse oximeter being tested after a period of time.

5. To print the results of the alarm, press the PRT softkey.

esl067.eps

The results will be sent to the RS232 or Centronics port.