Fluke Biomedical 2MF Index User Manual

Page 110

Advertising
background image

Index 2MF

Users Guide

F-2

Attachment of the
FingerClip Probe

Detector Window

Properly Applied

Finger Stop

Attaching the Probe to Index
(Ohmeda FingerClip Probe)

Misapplied

esl157.eps

Figure F-1. Attaching the Probe to Index 2MF

The low readings may be because the Pulse Oximeter under test is a Nellcor unit but

the user has Ohmeda selected. Make sure the make selected in Index 2MF agrees

with the make of the pulse oximeter being tested. Many manufacturers use different

“R” curves for measurement.
You might try repositioning the probe for closer results, especially disposable probes.

With disposable probes, it is difficult to ensure consistent alignment of the LEDs to

the photo detector. This could also be a problem when using a disposable probe on a

patient.

5. I cannot find the make of the pulse oximeter I am testing with Index 2MF. What do I

do?
Index 2MF currently has curves for many oximeters. If the pulse oximeter under test

is one of the following, use the Nellcor make selection:

Airshields, Inc.

Ivy Biomedical

Atom Medical Corp.

Marquette (Eagle)

Baxter Healthcare Corp.

MDE(Medical Data Electronics)

Century Medical Inc.

Mennen Medical

Colin Medical Instruments Corp.

North American Drager

Corometrics Medical Systems

Pace-Tech. Inc.

Critikon, Inc.

Protocol Systems, Inc.

Datascope Corp.

Pryon Corp

Drager (Europe)

Siemens

Edentec Sleeptrace/Areca

Science

Fukuda Denshi

SpaceLabs

Synectics Witt

Biomedical