Manufacturers Mode

Accessing Manufacturers' Tests

11

11-5

The Waveform Source screen provides two options:

esl146.eps

2. Select the oximeter by pressing UUT (Unit Under Test) as the source for the pulse

amplitude signal.

esl147.eps

3. Alternatively, press DC. The Simulator will ignore the oximeter's pulse amplitude

and will use instead a fixed dc level for infrared and red pulse simulations.

esl148.eps

The signals will pass through the Simulator based on their original amplitude.

4. To return to the Advanced Options menu, press esc.

Setting the R-Value

You can also enter a custom R-value using the following procedure:

1. Access the Advanced Options menu (if you have not already done so) by pressing

OPTS on Main Menu 2.

2. Press RVAL on the Advanced Options menu:

esl149.eps

You will see a screen appear similar to the following: