The MFE version ships with Nellcor and Ohmeda probe and oximeter
adapters. Please contact Fluke Biomedical regarding the availability of
additional adapter cables.

W

Caution

Attempting to create adapter cables without all details of the
oximeter probe-operating scheme could result in damage to the
oximeter, the Simulator, or both.

2. Press the

PRBE

softkey at Main Menu 1:

Fgg012.eps

The

PRBE

softkey initiates a testing sequence that includes:

Verifying that a Probe Test Board is installed in the Simulator

Checking whether or not a probe adapter cable is connected to the probe port

Performing all measurements for LED, photo detector, and resistance tests

esl125.eps

Probe Test Selection

The user can then choose the specific probe test.

LED and photodiode electrical test

Pin-to-pin resistances

Photodiode detector optical test

Note

During the photodiode test, the finger probe being tested should not be
attached to the Index finger.

LED Testing

The Probe test should end with the following menu displayed. Press the

LEDS

softkey:

Advertising