Printing – Fluke Biomedical 2MF Index User Manual

Page 93

Advertising
background image

Appendices

Printing and Data Transfer

A

A-5

Printing

Sample Index 2MF Test Results Output

The Simulator Autosequence printout shown below requires using the previous settings

to establish communication between the the Simulator and a printer.

The optional header is all information from the “Control#” field down to the

“Technician” field.

Actuals are all information under the “Settings:”

heading. A normal report includes

only the fields Alarm Response Time, Rate Alarm Response Time, and Pulse

Amplitude Test.

Fluke Biomedical
SPO2 Simulator Ver: 2.00
--------------------------------------
Control#: __________________________
Serial #: __________________________
Model: _____________________________
Mfr: _______________________________
Location: __________________________
Technician: ________________________
Date: ______________________________

Simulations: NELLCOR
---------------------------------------
Settings:

Actual:

#

O2

RATE

O2

RATE

1

100%

60 BPM

99 59

2

95%

65 BPM

95 64

3

90%

70 BPM

90 70

4

85%

75 BPM

85 75

5

84%

80 BPM

84 80

6

83%

85 BPM

83 85

7

82%

90 BPM

82 90

8

81%

95 BPM

81 95

9

80% 100 BPM

80 100

10

79% 105 BPM

79 105

---------------------------------------
Alarm Response Time:12 Seconds at 95% O2
---------------------------------------
Hi Rate Alarm Response Time:10 secs at 80
BPM
Lo Rate Alarm Response Time:10 secs at 80
BPM
---------------------------------------
Pulse Amplitude Test
At 85%, the signal is lost.
---------------------------------------
Motion Response Time: 10 Seconds
---------------------------------------
End of test.

esl155.eps