D. specifications, Specifications, Appendix d – Fluke Biomedical 2MF Index User Manual

Page 99: Equipment specifications

Advertising
background image

D-1

Appendix D

Specifications

Equipment Specifications

The following specifications apply to the Index 2MF and 2MFE.

%O2

Range ................................................................. 35 to 100 %
Resolution........................................................... 1 %

Accuracy ............................................................. 100 to 75 %: +1 %

±

accuracy of the pulse oximeter under test.

74 to 50 %: + 3 %

±

the accuracy of the pulse oximeter under test.

<50 %: unspecified

Repeatability ....................................................... +1 standard deviation

Rate

Range ................................................................. 30 to 250 BPM
Resolution........................................................... 1 BPM
Accuracy ............................................................. 1 % +1 BPM

Pulse Amplitude

Range ................................................................. 0 to 20 %
Resolution........................................................... 1 %

Battery Life ............................................................ At least 4 hours of continuous use
Battery Charger

Type .................................................................... Lead-Acid battery charger only
Input .................................................................... 100 – 250 V, 60 – 60 Hz, 0.3 A
Output ................................................................. 12 V dc, 0.5 A

Probe Test ............................................................. Index 2MFE version only
Continuity/Resistance Test

Test Matrix .......................................................... Measures all combinations of possible interconnections in an XX point

matrix.

Range ................................................................. 250

Ω

to 150 k

Ω

Accuracy ............................................................. – 5 % of reading

LED/Detector Voltage Test

Test Format ........................................................ Measures the voltage drop across Red LED, infrared LED, and the

photo detector when the internally generated test signal is applied.

Test Signal.......................................................... Constant current source at 1.0 mA
Open Circuit........................................................ 2.5 V max
Measurement/Display Range ............................. 0.0 to 4 V
Accuracy ............................................................. – 5 % of reading, 0.4 to 4 V