Chapter 11

Manufacturers Mode

Introduction

Fluke Biomedical has included several features in the Simulator specifically designed to
aid manufacturers of pulse oximeters. Those features ease the following manufacturing-
specific needs:

The automatic production testing of pulse oximeters

The automatic production testing of oximeter probes

The design and development of new pulse oximeters

Automated Testing

With the Simulator, manufacturers of pulse oximeters can test their products using a
standard PC. The computer must have a serial communications (COM) port and a
communications program, such as the Microsoft Windows

TM

“HyperTerminal” program.

To test a pulse oximeter using a PC, connect the Simulator to the computer via one of the
communications ports, with the unit under test connected via a second communications
port.

You send special commands to the Simulator using the computer; the Simulator then acts
on the commands to test an oximeter. You can additionally use the computer to monitor
responses of the unit under test (UUT). With the computer, you can do any of the
following:

Log the oximeter's responses for a production test record

Compare the oximeter's responses to preset limits to determine whether the

system passes or fails quality standards

You can complete these procedures by testing oximeters against a reference probe.
Alternatively, you can test the probes against a reference oximeter.

Manufacturer Parameters Available For Simulations

Although percent oxygen and beats per minute are sufficient for a basic check of an
oximeter, these simulations cannot by themselves probe the limits of an oximeter's
performance in order to evaluate the its design.