7.2 Reports

96

780/781 pH/Ion Meter, Manual

7.2

Reports

A pH/Ion Meter report is produced by transmitting data to a printer or

PC via the RS232 interface. In this way you can document e.g. measur-

ing results, calibration data, method parameters or the configuration of

the instrument in written form.
A requirement for a properly functioning report output is the input of the

receiving instrument (see Section 5.6) and the correct setting of the

RS232 transmission parameters (see Section 5.7).
The following reports can be automatically printed out after a meas-

urement has been made: a measuring point report is produced directly

with the <MEAS/PRINT> key. The automatic output of calibration,

electrode test and result reports must be defined as such under meth-

od parameters.
Apart from the measuring point report, all reports can be triggered

manually at a later date by the <REPORT> key. A dialog appears for

the selection of the required report with the <SELECT> keys:

user memory

Report of the memory occupancy by methods and

calibration data

calib short

Calibration report without graphical calibration

function curve

calib full

Calibration report with graphical calibration func-

tion curve

config

Instrument configuration report

param

Current method parameters report

result short

Measuring result report for standard or sample

addition without graphical presentation

(781 pH/Ion Meter only)

result full

Measuring result report for standard or sample

addition with graphical presentation

(781 pH/Ion Meter only)

el.test

pH electrode test report (see Section 8.6)

mv memory

Output of all stored measured values

all

Output of all existing reports

During selection these reports can also be selected directly by using

the key which has the corresponding function (e.g. <CAL.DATA> for a

calibration report).