6.4 Potential measurement (U mode)

70

780/781 pH/Ion Meter, Manual

state:

ON, OFF

u.limit

-2200.0...1000.0...2200.0 mV

u.hyst.

-2200.0...2.0...2200.0 mV

l.limit

-2200.0...-1000.0...2200.0 mV

l.hyst.

-2200.0...2.0...2200.0 mV

The exact way in which the limits function works is described in Sec-

tion 7.4 independent of the mode.

6.4.3

Plot parameters

PARAMETER U

:

plot parameters

: ├

left mar.

right mar.

The measured potential can be printed out as a function of time in a

plot. Under

CONFIG/print meas.value

the print criterion '

plot

'

must be selected (see Section 5.2). The scales of the measured values

axes are determined by the following limits.

Left mar.

-2200.0...-1000.0...2200.0 mV

Right mar.

-2200.0...1000.0...2200.0 mV

TIP !

The value for the right-hand limit of a plot must not necessarily be the

larger one. By exchanging the limits the plot can be mirrored.

6.4.4

Preselections

PARAMETER U

:

preselections

req.ident

req.ident:

id1, id1 & id2, OFF

For sample identification you can document each measured value with

two Ids in the report header. These can either be predefined under

CONFIG/report

and included in each report or, depending on the Pre-

selection setting, they can be requested before each measurement.
Please note that sample identifications requested as a preselection be-

fore the measurement will also be stored as an entry under

CON-

FIG/report

. This means that they are again available for the next

measurement.