2 ph measurement (ph mode), 1 measuring parameters, Ph measurement (ph mode) – Metrohm 781 pH/Ion Meter User Manual

Page 66: Measuring parameters

Advertising
background image

6.2 pH measurement (pH mode)

56

780/781 pH/Ion Meter, Manual

6.2

pH measurement (pH mode)

6.2.1

Measuring parameters

PARAMETER pH

measuring parameters

: ├

electr.id

drift pH

temperature*

method

delta measurement

stirrer

(* only appears if no temperature sensor is connected)
The measuring parameters describe the settings which are of direct

importance for the measurement.

electr.id:

12 ASCII characters

You can assign any name you like to the connected pH electrode. Dur-

ing a calibration the resulting calibration data will be assigned to this

entered electrode Id and stored. The electrode Id will then be included

in the report header for the output of measured values.
In this way you can enter several electrode Ids. This allows you to

switch between various pH electrodes with their associated calibration

data. However, a requirement is that each new electrode Id is linked

with a new calibration. Then you can use <SELECT> under the meas-

uring parameter

Electr.Id

to select one of the stored electrode Ids.

In order to delete an electrode Id and its linked calibration data this

must be reset under <CAL.DATA> (

Reset cal.

, see Section 7.1.1).

drift pH

0.001...0.050...9.999/min, OFF

A measuring signal which changes as a function of time is regarded as

being sufficiently constant when its drift drops below the value defined

here. This can be used as criterion for the pH/Ion Meter if a displayed

measured value is to be stored or documented by printing a report (see

print criterion and storage criterion in Section 5.2 and Section 5.3).

Please note that for a drift-controlled measured value recording this drift

parameter must not be set to '

OFF

', as otherwise no result will be ex-

ported.
If the drift condition is not fulfilled then this is indicated by two constant-

ly blinking triangles in the display.
An active drift control increases the reproducibility of the measurements

as a constant condition is used to define the measured result. A smaller

value for the measuring parameter Drift pH requires a more stable

measuring signal than a larger one. The set measuring parameter is al-

so proportional to the measuring uncertainty of the measured value ob-

tained.

Advertising
This manual is related to the following products: