5 Configuration

780/781 pH/Ion Meter, Manual

45

Normally each measured value is accompanied by the report

header for documentation purposes. The exact contents of the

report header are defined under CONFIG/Report (see above).

With a larger series of manually triggered measurements it may

be advisable to print out the report header only once with the first

measured value. In this case all the measurements should be

made under the same conditions. The report header will only ap-

pear again when the run number has been reset to zero by

switching the instrument off and on.

calibration report: ON, OFF

You have the possibility of providing each measured value with

the associated calibration data of the electrode used, e.g. within

the context of a GLP documentation.

print crit:

time

With this print criterion you can automatically record and output the

measured values at fixed intervals. The following options are used:

date & time:

ON, OFF

report header:

once, always, OFF

calibration report: ON, OFF
These settings correspond to those of the print criterion 'immedi-

ate' (see above).

time interval:

0.1...4.0...99999.0 s

Automatic measuring value recording can be carried out at time

intervals of down to 0.1 s. Please note that if small time intervals

are used then the amount of data for each measured value must

be adapted accordingly, as otherwise the buffer memory of the

connected printer or PC would be filled too quickly. This means,

for example, that the report header should only be printed out

once at the start of a series of measurements. Otherwise data

could be lost at the RS232 interface of the receiver. In this case

you should consider matching the measured value recording in-

terval to the measured value memory (see Section 5.3).

stop time:

OFF, 1...999999 s

The time after which an automatic series of measurements is to

be terminated can be entered in whole seconds. You can also let

a series of measurements run indefinitely without a fixed stop time

and then stop it manually with <QUIT> if you select 'OFF' for this

setting.