3 ph calibration, Ph calibration – Metrohm 781 pH/Ion Meter User Manual

Page 30

Advertising
background image

3.1 pH measurement

20

780/781 pH/Ion Meter, Manual

2 Initial configuration

Switch on pH/Ion Meter

Section 5.4

Set dialog language

Section 5.4

Set date/time

Section 5.4

Set temperature sensor

Section 5.4

3.1.3

pH calibration

1 Start calibration with first buffer

Immerse pH electrode in buffer solution pH 7

Switch on stirrer with <STIRRER>

Start calibration with <CAL>

If no temperature sensor is connected:

enter the temperature and exit with <ENTER>

2 Continue calibration with second buffer

Remove pH electrode from first buffer, rinse with water and

blot off gently with a lint-free tissue

Immerse pH electrode in second buffer solution pH 4 and

continue calibration sequence with <ENTER>

3 Result

Wait for potential measurement, result will be displayed:

After 30 s the pH/Ion Meter will switch back automatically to

the measurement display. This can also be done immediately

with <ENTER>.

CONFIG

CAL

ON/OFF

ENTER

ENTER