7.2 Reports

104

780/781 pH/Ion Meter, Manual

The settings shown for the instrument configuration correspond to the

standard settings in the normal operating condition of the pH/Ion Meter

after the configuration memory has been initialized (see Section 8.5) fol-

lowed by conversion to English as the dialog language.

7.2.7

Parameter report

in order to document the parameter settings for the current method you

can produce a parameter report:

'pa

781 pH/Ion Meter

01104 5.781.0020

date 2004-06-03 12:08:36

PARAMETER pH

>measuring parameters

electr.id

drift 0.050 /min

method ********

delta measurement: OFF

stirrer: OFF

>calibration parameters

drift 0.5 mV/min

report: OFF

cal.interval OFF h

no.of buffers 2

buffer type: Metrohm

low lim.slope 95.00 %

up lim.slope 103.0 %

low lim.pH(0) 6.40

up lim.pH(0) 8.00

offset Uoff state: OFF

>limits pH

state: OFF

>limits T

state: OFF

>plot parameters

left mar.: 0.0

right mar.: 14.0

left marg.T 20 °C

right marg.T 30 °C

>preselections

req.ident: OFF

>electrode test

electrode type: standard

report: OFF

--------------

The standard parameters following a memory initialization in the current

mode are shown; in this case the pH mode with a connected tempera-

ture sensor (see Section 8.5). Such parameters can be stored as a

method (see Section 6.1).

PARAM

REPORT

9

+