6 Methods / Parameters

780/781 pH/Ion Meter, Manual

57

Please note that the drift condition defined here and the temperature

parameter described below do not apply to the calibration and, if nec-

essary, must be set separately under the calibration parameters.

temperature

-999.9...25.0...999.9 °C

This menu item only appears when no temperature sensor is connect-

ed. In this case you can enter the temperature of the sample solution

here. If the measuring and calibration temperatures are different then it

is necessary to record the temperature for the automatic correction of

the electrode slope (temperature compensation, see Section 9.2.1). In-

formation about the measuring temperature is also absolutely neces-

sary for the complete documentation of a pH value. In the report

printout of the measured value this manually determined temperature is

indicated by the suffix '

manual

'.

method

8 ASCII characters

All parameter settings can be stored together or loaded as a method

(see Section 6.1). The name of the current method is then given here.

This field is only used for information here and cannot be edited direct-

ly.

delta measurement ON, OFF
The delta measurement permits the display of the measured value rela-

tive to a freely selectable constant reference value:

reference pH

-19.999...0.000...19.999

This reference value is subtracted from the measured value itself. The

modified measured value is shown together with this value and is taken

into consideration, for example, in checking the limit values (see

PA-

RAMETER pH/limits pH

). The reference value is not included in the

report. However, should you still want to document it then you should

also produce a parameter report (see Section 7.2.7).

stirrer:

ON, OFF, control

You can choose whether the pH measurement is to be made while stir-

ring is being carried out or not. A requirement is that a Metrohm stirrer

is connected with a suitable cable (see Section 2.3). If an 8xx series stir-

rer has been connected and the settings made under

CON-

FIG/peripheral units

then the stirring rate can also be selected:

stirring rate

1...5...15

If you want your measurement to be made without stirring but still want

to stir the solution before the measurement then you should select the

control

setting. In this way you can trigger a stirring sequence before

the start of each measurement with <MEAS/PRINT>:

prestir pause

0...99999 s

stir time

0...99999 s

poststir pause

0...99999 s