8.5 Diagnosis

124

780/781 pH/Ion Meter, Manual

8.5

Diagnosis

The pH/Ion Meter is equipped with a diagnosis program. This allows

selective checks of the functions of individual assemblies.
The diagnosis can be carried out directly after the instrument has been

switched on by pressing and holding down the <9> key. Only those

tests that you can carry out yourself are described below; the other

functions are reserved for Metrohm Service personnel.
If one of the following tests is concluded with an error message, please

note it down and contact your local Metrohm agency.

You are expressly warned against triggering any other functions other

than those listed and documented below. Improper use can disturb

the correct functioning of the pH/Ion Meter and may render it unusa-

ble.

diagnose

RAM Initialization

RAM Test

AD-Converter Test

LCD Display Test

:

Key Test

:

RAM Initialization

Select: ACTMODE, MODES, SETUP, CONFIG,

ASSEMBLY, ALL

The initialization of the instrument memory can be a good idea if you

want to return parameter and configuration settings to their original

condition. This does not affect the instrument software itself, but only

the above selection of the corresponding user data. ACTMODE resets all

method parameters of the current measuring mode, MODES affects the

method parameters of all modes. With SETUP you can initialize the set-

up settings, with CONFIG all the configuration data will be reset. AS-

SEMBLY functions, which can only be accessed via remote control, are

reset here. The option ALL resets the whole instrument memory to its

original condition.

A memory initialization should only be triggered after careful consider-

ation! It will permanently delete all the affected user data which must

then be reentered manually.

When only certain configuration settings or method parameters are to

be reset to their default values then the manual version is more suita-

ble. Except for a few special cases, all the corresponding default values

can be set directly during input with the <CLEAR> key. In the other

cases you will find the default settings in this Manual in Sections 5 and

6.