7.2 Reports

100

780/781 pH/Ion Meter, Manual

If a whole series of measured values is to be documented then this is

done by selecting under

CONFIG/Print meas. values

the option

Report header: once

. In this way the report header will be printed

out once with the first measured value. Each further time that the

<MEAS/PRINT> key is pressed a further measured value will be print-

ed out:

#2 pH = 8.185 21.2 °C (Pt1000)

2004-05-28 17:01:56

#3 pH = 8.187 21.2 °C (Pt1000)

2004-05-28 17:02:56

#4 pH = 8.188 21.2 °C (Pt1000)

2004-05-28 17:03:56

#5 pH = 8.189 21.2 °C (Pt1000)

2004-05-28 17:04:56

Such a series of measured values can also be produced automatically

by selecting the print criterion '

time

'.

With the print criterion 'plot' the measured values are given out as a

graphical plot in the direct measurement types. As well as configuration

under

CONFIG/print meas.value

, the scales of the measured values

curves must be set under

PARAMETER/plot parameters

(see

plot

parameters

in Section 6.2 to Section 6.5). In the pH and Conc(direct)

modes the measured temperature can also be shown in the same plot

as a secondary measured value in addition to the primary measured

value, pH or concentration:

'mp

781 pH/Ion Meter

01104 5.781.0020

date 2004-05-28 17:00:56

user

C. Weber

pH pH-1 run number

9

electr.id

pH electr.01

Temperature scale

pH scale

Temperature curve

pH curve

lower limit pH

upper limit pH