8.3

QM support

Metrohm AG can support you in your quality management (QM)

measures:

Literature
Brochures in the series "Quality management with Metrohm" provide an

overview of the topics of quality assurance and validation, help with the

documentation of your validation activities and explain the quality man-

agement system at Metrohm AG, Herisau. You can request them under

www.metrohm.com or from your local Metrohm agency.
Further information about the validation of Metrohm pH meters can be

found in Application Bulletin No. 271.

Instrument-internal validation aids
The system test of the pH/Ion Meter checks that all the relevant assem-

blies are functioning properly when the instrument is switched on. The

result can be produced directly as a system test report via the RS232

interface (see Section 5.5).
Calibration is a simple and necessary means of checking the functions

of the pH/Ion Meter and the electrodes connected to it. The determined

calibration data is checked according to certain criteria, can be as-

signed to a defined electrode identification and can be stored together

with it. When a defined calibration interval has expired the instrument

will inform you that the recalibration of the connected electrode is nec-

essary. A description of the calibration parameters for the various

modes can be found in Section 6.2.2 (pH) and Section 6.5.5 (Conc [781

only]).
You can use monitoring to have the instrument remind you automatical-

ly about tasks which are due for validation purposes or service

measures. Just as for an expired calibration interval, these messages

will also be protocolled with each output of a measured value which has

been determined since the expiry of the due date even though the nec-

essary measures have not yet been taken.
With diagnose you can subject the various assemblies of the pH/Ion

Meter to further-going tests. This can help you to localize any faults

which may occur. This function can also be integrated in your validation

measures. It is described in detail in Section 8.5.
The electrode test is a programmed sequence of various measure-

ments for checking the quality of the connected pH electrode. Sophisti-

cated assessments help with the selective localization of an appearing

problem. In Section 8.6 you will find an exact description of how to carry

out such an electrode test, a detailed interpretation of the possible test

results as well as the appropriate measures for remedying the faults

and preventing future ones.