8 Troubleshooting – Messages – Maintenance

780/781 pH/Ion Meter, Manual

131

8.6.4

Messages and measures

For the test results '

passing electrode

' and '

electrode out of

limits

' the following recommendations for improving the quality of the

electrode can be given:

Result

Measure

passing electrode

Clean diaphragm

electrode out of limits

Clean diaphragm and/or re-

generate the glass membrane;

check reference system

The following general information should be taken into consideration

when evaluating the electrode test results:

With separate measuring and reference electrodes the pH elec-

trode is not always the cause of unexpectedly poor test results.

This is why you should also check the condition of the reference

electrode (see Section 8.6.5).

If the tested pH electrode has been given a lower evaluation only

because of its slope being inadequate then you should compare

the individual slopes at pH 4/7 and pH 7/9. If these values are

different then repeat the electrode test using a fresh pH 9 buffer.

Only then can you conclude that there is an electrode problem.

A correct outflow of the inner electrolyte causes only a minimum

potential drop when the stirrer is turned of (see Fig. 14). A large

drop indicates a too small outflow - the diaphragm is blocked. If

you obtain a potential step-up, the outflow is too high and should

be corrected.

Under certain conditions, e.g. very low relative humidity, plastic

floor coverings or clothing made from synthetic materials, the pH

electrode can be affected by electrostatic discharges. The re-

sults are high drift values and correspondingly poor electrode

test results. Such problems can be avoided by the user wearing

e.g. an antistatic bracelet.

In addition to the evaluation of the electrode quality described above,

the measured values are also checked to see if they have exceeded

certain tolerances. In such cases the electrode test is concluded with

the following message on the display:

electrode test not passed

More specific error messages will also be shown and included in the

report. The following table lists these tests, the resulting messages and

recommended measures for remedying their possible causes. These

measures are described in detail in the following Section.
The first two tests are carried out after the measurement in the pH 9

buffer and the electrode test then stopped if necessary.