7 Various functions

780/781 pH/Ion Meter, Manual

85

7

Various functions

This section describes various functions of the pH/Ion Meter which can-

not be assigned to individual modes.

7.1

Calibration and addition/subtraction data

The current pH calibration data can be called up at any time with the

<CAL.DATA> key. For the 781 pH/Ion Meter the same applies to the

concentration calibration (<CAL.DATA>) and the addition or subtrac-

tion data (<ADD.DATA>) in the Conc mode.
A detailed description of the evaluation of the calibration and addi-

tion/subtraction measurements for the calculation of this data can be

found in Section 9.2.
An overview appears which, depending on the measuring mode and

calibration data, can have a different composition.
For example, if no calibration has yet been carried out for a new elec-

trode Id then only the theoretical calibration data slope = 100.00% and

pH(0) = 7.000 will be shown and taken into account for the measure-

ment.

7.1.1

pH calibration data

CAL.DATA

7

CAL.DATA

electr.id

method

slope

pH(0)

temperature

temp.sensor

c.dat.

cal.interval

variance

buffer type

no.of buffers

cal.table

b1

b2

:

bn

curve

electr.id
If an electrode Id has been selected or entered under

PARAMETER pH/

measuring parameters/electr.id

before the calibration then it will

be shown here. The calibration data apply only to this electrode and will

change when a new electrode Id is set under the above measuring pa-

rameters (see Section 6.2.2).