9.2 Evaluation

140

780/781 pH/Ion Meter, Manual

The calculation of the calibration data depends on the number of pH

buffers used:
1-Point calibration
Slope:

The theoretical slope value of 100.0% is used.

pH(0)

=

N

1

1

U

Slope

U

pH

2-Point calibration

Slope

=

(

)

(

)

N

1

2

1

2

U

1

pH

pH

U

U

pH(0)

=

N

U

Slope

U

pH

;pH , U = mean value

3- and more point calibration
With at least 3 calibration buffers a compensation curve according to

the principles of the smallest mean-square error (linear regression) will

be calculated.

Slope

=

(

) (

)

(

)

N

i

2

i

i

i

i

U

1

pH

pH

U

U

pH

pH

pH(0)

=

N

U

Slope

U

pH

Variance =

(

)

2

n

U

U

i

2

calculated

,i

i

;n = no. of measuring points