2.3.3

Connecting a Dosimat Plus (only 781)

Controlled addition of standard or sample solutions
For automatic calibration as well as for the addition of standard or

samples in the Conc mode with the 781 pH/Ion Meter the Metrohm 865

and 876 Dosimat Plus devices can be connected via the separately

available 6.2148.010 Remote Box. Control is carried out via the

6.2141.350 cable:

Fig. 9: 781 pH/Ion Meter – Dosimat Plus

If a Dosimat Plus is used then a stirrer is normally also required. This is

then connected directly with the MSB connection of the 6.2148.010

Remote Box (stirrer: 801 or 802+804).

Please note that for the standard addition the mode “

DOS

” is set on

the Dosimat Plus. The activation pulse must be activated on the pH

Meter as well as the Dosimat Plus. Furthermore, the Dosimat Plus

must be in the status

“HOLD”

. You can set the status to

“HOLD”

by

pressing the “Go” key once (check: the display of the Dosimat Plus

shows

“HOLD”

in the top right-hand corner).