8.2 Messages

118

780/781 pH/Ion Meter, Manual

Message

Situation

Cause

Remedy

buffers
unsuitable

Electrode test

Electrode test has been
started with a set of buffers
that does not contain all
the necessary buffers

Change buffers

Check selected buffer

type in the calibration pa-
rameters (see Sec-
tion
8.6.1)

change buffer
<ENTER>

pH calibration

Instrument is waiting for
new buffer

1. Change buffer
2. Press

<ENTER>

buffer alloc.
impossible?

pH calibration

Buffer not recognized or

not defined (buffer type
'special', 'own' or 'mixed')




Offset potential Uoff is

set too high

1. Replace buffer
2. Press

<ENTER>

or
1. Stop calibration
2. Define buffer
3. Restart calibration
or
Enter Uoff (see

Sec-

tion 6.2.2

) and restart cali-

bration

report output
...

Standard addi-
tion,
sample addition

Instrument is busy with the
output of a RS232 report

Wait until report output is
finished

RS error 36

RS232 remote
control

RS receiver error:
parity unmatched

1. Press

<QUIT>

2. Match the parities of both

instruments (see

Sec-

tion 5.7

)

RS error 37

RS232 remote
control

RS receiver error:
stop bit unmatched

1. Press

<QUIT>

2. Match the stop bits of

both instruments (see

Section 5.7

)

RS error 38

RS232 remote
control

RS receiver error:
data overflow, at least one
character cannot be read
(incorrect baud rate)

1. Press

<QUIT>

2. Match the baud rates of

both instruments (see

Section 5.7

)

RS error 39

RS232 remote
control

RS receiver error:
internal receiver buffer
overflow

1. Press

<QUIT>

2. Repeat process

RS error 42

Report output

RS transmitter error:
handshake not replied to
for at least one second

1. Press

<QUIT>

2. Check reception readi-

ness at other end

RS error 43

Report output

RS transmitter error:
transmission from the pH
Meter has been interrupted
by XOFF for at least three
seconds

Press <QUIT>

Send XON