4 ph measurement, Ph measurement – Metrohm 781 pH/Ion Meter User Manual

Page 32

Advertising
background image

3.1 pH measurement

22

780/781 pH/Ion Meter, Manual

3.1.4

pH measurement

1 Select printout criterion

If the obtained measured value is to be printed out directly as

a measuring point report or transferred to a PC then the re-

quired printout criterion must be set

(see Section 5.2):

CONFIG/print meas.value/print crit.: drift

2 Start measurement

Immerse the pH electrode in the sample.

Switch on stirrer with <STIRRER>.

3 Determine measured value

You can read off the actual pH of the sample from the display.

It is stable when the drift display

(small double arrow on the left-hand side) no longer blinks.

4 Print measurement

With <MEAS/PRINT> you can start the output of a measuring

point report on a printer or PC. The recording of the meas-

urement and its output depend on the printout criterion 'Drift'

and take place only when the current drift condition is fulfilled

(see Section 6.2.1):




CONFIG

STIRRER

2

MEAS/

PRINT

'mp
781 pH/Ion Meter 01104 5.781.0020
date 2004-05-12 16:57:55
pH ******** run number 5
electr.id
#5 pH = 5.875 20.2 °C Pt1000
==============