6 Methods / Parameters

780/781 pH/Ion Meter, Manual

63

6.2.5

Plot parameters

PARAMETER pH

:

plot parameters

: ├

left mar. pH

right mar. pH

left marg.T

right marg.T

The two directly recorded measuring quantities pH and temperature

can be printed out as a function of time in a plot. Under

CONFIG/print

meas. value

the print criterion '

plot

' must be selected (see Sec-

tion 5.2). The scales of the measured values axes are determined by

the following limits.

left mar. pH

-19.9...0.0...19.9

right mar. pH

-19.9...14.0...19.9

left marg.T

-999...20...999 °C

right marg.T

-999...30...999 °C

TIP !

The value for the right-hand limit of a plot must not necessarily be the

larger one. By exchanging the limits the plot of the particular quantity

can be mirrored.

6.2.6

Preselections

PARAMETER pH

:

preselections

: └

req.ident

req.ident.:

id1, id1 & id2, OFF

For sample identification you can document each measured value with

two Ids in the report header. These can either be predefined under

CONFIG/report

and included in each report or, depending on the pre-

selections setting, they can be requested before each measurement.
Please note that sample identifications requested as a preselection be-

fore the measurement will also be stored as an entry under

CON-

FIG/report

. This means that they are again available for the next

measurement.