7.4 Limit monitoring

110

780/781 pH/Ion Meter, Manual

7.4.2

How it functions

The exact way in which limit monitoring functions and the effect on the

remote outputs are shown in the following diagram.

time

measured

value

u. limit

u. hysteresis

l. hysteresis

signal output

l. limit

u. limit

l. limit

aktive

inaktive

aktive

inaktive

Fig. 13: Remote outputs in limit monitoring

The lower and upper limits and the associated hysteresis are the char-

acteristic quantities for limit monitoring. They can be defined in the pa-

rameter settings for each measuring mode (see Section 6.2 to 6.5). The

lower or upper limit is always the triggering point for the activation of the

corresponding remote line or for the message on the display. If the up-

per limit is again undercut or the lower limit exceeded again then the

activation will continue until the so-called hysteresis range has also

been exited. This avoids the too frequent activation or deactivation of

the alarm function when the measured value oscillates very closely

around a limit. This means that the activation cancellation point is de-

fined as the upper limit minus hysteresis and the lower limit plus hyste-

resis.
Set the hysteresis to zero, if you are just interested in an exact docu-

mentation of limit overruns.