2 preselections, Preselections – Metrohm 781 pH/Ion Meter User Manual

Page 93

Advertising
background image

6 Methods / Parameters

780/781 pH/Ion Meter, Manual

83

auto dos

increment 1...n 1.0E-30...0.1...999.9 mL
The particular volume increment must be entered for each addi-

tion of the standard or sample solution.

auto

delta U

1...10...999 mV

Under

auto

the required potential difference must be entered.

Please note that the Dosimat Plus must be properly prepared be-

fore the start of an addition run (cylinder freshly filled, air bubbles

removed, etc.).

During an addition you can terminate the run prematurely with it

<MODE>; only the additions which have already been measured will

be used for the calculation of the result. This manual termination will be

marked in the report.

6.6.2

Preselections

PARAMETER Conc

:

preselections

req.ident

req.smpl.size

activate pulse

req.ident:

Id1, Id1 & Id2, OFF

For sample identification you can document each measured value with

two Ids in the report header. These can either be predefined under

CONFIG/Report

and included in each report or, depending on the Pre-

selections setting, they can be requested before each measurement.
Please note that sample identifications requested as a preselection be-

fore the measurement will also be stored as an entry under

CON-

FIG/Report

. This means that they are again available for the next

measurement.

req.smpl.size : value, unit, all, OFF
The sample size and sample size unit parameters described under the

calculation parameters (see Section 6.5.4) can also be requested be-

fore each measurement. This allows a more individual input of the

sample data.