3 temperature measurement (t mode), 1 measuring parameters, Temperature measurement (t mode) – Metrohm 781 pH/Ion Meter User Manual

Page 75: Measuring parameters

Advertising
background image

6 Methods / Parameters

780/781 pH/Ion Meter, Manual

65

6.3

Temperature measurement (T mode)

6.3.1

Measuring parameters

PARAMETER T

measuring parameters

: ├

electr.id

drift

method

delta measurement

stirrer

The measuring parameters describe the settings of direct importance

for the measurement.

electr.id

12 ASCII characters

You can assign any name you want to the connected temperature sen-

sor. This sensor Id is then contained in the report header for measured

values.

drift

0.5...1.0...999.9 °C/min, OFF

A measuring signal which changes as a function of time is regarded as

being sufficiently constant when its drift drops below the value defined

here. This can be used as criterion for the pH/Ion Meter if a displayed

measured value is to be stored or documented by printing a report (see

print criterion and storage criterion in Section 5.2 and Section 5.3).

Please note that for a drift-controlled measured value recording this drift

parameter must not be set to '

OFF

', as otherwise no result will be ex-

ported.
If this drift condition is not fulfilled then this is indicated by two constant-

ly blinking triangles in the display.
An active drift control increases the reproducibility of the measurements

as a constant condition is used to define the measured result. A smaller

value for the measuring parameter Drift requires a more stable measur-

ing signal than a larger one. The set measuring parameter is also pro-

portional to the measuring uncertainty of the measured value obtained.

method

8 ASCII characters

All parameter settings can be stored together or loaded as a method

(see Section 6.1). The name of the current method is then given here.

This field is only used for information here and cannot be edited direct-

ly.

delta measurement: ON, OFF
The delta measurement permits the display of the measured value rela-

tive to a freely selectable constant reference value:

reference

-999.9...0.0...999.9 °C

Advertising
This manual is related to the following products: