8.6 pH electrode test

128

780/781 pH/Ion Meter, Manual

1 Start pH electrode test

The electrode test is started from the pH measured value dis-

play with <EL.TEST>.

2 (Enter temperature)

If no temperature sensor is connected then the measuring

temperature of the buffers must be entered here. This should

be the same for all buffers.

3 Immerse electrode in pH 9 buffer

Immerse the electrode in the pH 9 buffer and immediately

press <ENTER>.

The instrument now counts down from 239 to 0 seconds.

If a Metrohm stirrer is connected and configured (see Sec-

tion 8.6.1) then it will be switched off automatically after 3

minutes. Otherwise you will be requested to switch off the stir-

rer; this request must be confirmed with <ENTER>.

If a temperature sensor is connected the temperature meas-

urement will then be carried out.

4 Immerse electrode in pH 4 buffer

Immerse the electrode in the pH 4 buffer and immediately

press <ENTER>.

Further procedure as for the pH 9 buffer.

5 Immerse electrode in pH 7 buffer

Immerse the electrode in the pH 7 buffer and immediately

press <ENTER>.

Further procedure as for the pH 9 buffer.

6 Evaluation of the pH electrode and report output

The electrode evaluation will then appear on the display. If ap-

plicable the report will also be printed out on a connected

printer.

If interference occurs the electrode test may be terminated.

Such interference is also shown on the display and noted in

the report (see Section 8.6.4)