Metrohm 781 pH/Ion Meter User Manual

Page 56

Advertising
background image

5.2 Printing out measured values

46

780/781 pH/Ion Meter, Manual

print crit.: drift
If this print criterion is used then a measured value will only be docu-

mented when the drift condition defined under

PARAMETERS/measur-

ing parameters/drift

has been fulfilled.

date & time:

ON, OFF

report header:

once, always, OFF

calibration report: ON, OFF
These settings correspond to those of the print criterion

'

immediate

' (see above).

print crit.: plot
The directly recorded values can be outputted graphically as a meas-

ured value/ time plot.

time interval:

0.1...4...99999.0 s

In a similar way to the print criterion 'time' a time interval can be

defined for graphical measured value recording.

time scale:

5,10,30,60,120,180..99960

s/cm

This is the time axis scale. The reciprocal value of the quantity en-

tered corresponds to the paper feed rate. This means that, e.g.,

for a required feed rate of 2 cm/min a time scale of ½ min/cm =

30 s/cm must be entered.

time scale label.: rel, abs

The time axis scale can use the relative time in minutes and sec-

onds (start = 0 s). The current time can also be used with '

abs

'.

stop time:

OFF, 1...999999 s

The graphical plot is terminated when the stop time has elapsed

or will continue indefinitely ('

OFF

'). In both cases termination is

possible with <QUIT>.

Advertising