6 Methods / Parameters

780/781 pH/Ion Meter, Manual

81

type:

add, sub

In general – and also in this Manual – when dealing with standard and

sample addition techniques it is addition which is described. It is also

possible to carry out a standard or sample subtraction, for example by

precipitation reactions. The appropriate type is set here.

conc.std.

1.0E-30...1.0...1.0E+30

The concentration of the standard solution used is to be entered here,

regardless of whether it is to be added during standard addition or al-

ready present for sample addition.

report:

OFF, short, full, line

After the addition run you can automatically output the result in the form

of report. You can choose between the '

short

' version containing all the

relevant data and the '

full

' version, which also shows the measuring

curves as a graph (see Section 7.2.5). The '

line

' setting supplies re-

ports with the most important data in the same form as in the measur-

ing point reports for direct measurements.
The '

short

' and '

full

' reports can also be produced later under

<REPORT>.

addition:

manual, auto dos, auto

Standard or sample addition can be carried out in various ways:

manual
The standard or sample solution is added manually at defined

volume increments.

no.of additions 1, 2, 3,...19
The number of additions is entered here. If only one addition is

made then the program will adopt for the slope the existing value

from the current calibration data of the electrode used. This will

also be corrected accordingly if measuring temperature changes

in comparison to the calibration temperature. This adoption of the

existing slope also applies when a standard or sample addition

with several additions is canceled with <MODE> or <QUIT> af-

ter the first addition and then confirmed with <ENTER>.

However, please note that, as for the calibration of ion-selective

electrodes, at least two additions should be made (three to five or

even more would be better). The possibility of terminating the ad-

dition after a single standard or sample addition as described

above should only represent an absolute exception.

increment 1...n 1.0E-30...0.1...999.9 mL
The volume increment for each addition is entered here.